Method Development and Validation for the Determination of Pravastatin in Human Plasma by Lc-Ms/Ms

[ ISSN : 2573-3729 ]

Abstract Citation Introduction Materials and Methods Results Discussion Conclusion References
Details

Received: 22-May-2017

Accepted: 06-Jun-2017

Published: 09-Jun-2017

Lian Chen, Paresh Joshi, Andrii Piatkivskyi, Kalem Aguilar, and Jenny Lin*

Department of Pharmaceuticals, CMIC Inc, United States

Corresponding Author:

Lian Chen, CMIC, Inc. 2860 Forbs Avenue, Hoffman Estates, IL, 60192, United States, Tel: +847-645-0407 (Ext 139); Fax: 847-645-0412; Email: lian-chen@cmicgroup.com

Abstract

A simple, rapid, sensitive and selective liquid chromatography/tandem mass spectrometry (LC/MS/MS) method was developed and validated for the quantification of pravastatin in human plasma. Pravastatin-D3 was used as an internal standard. The analyte was extracted from human plasma samples by liquid-liquid extraction technique. Due to the presence of isobaric metabolites, 3α-iso-pravastatin and 6-epi-pravastatin, chromatographic conditions were optimized, with a C18 column by using a mixture of 0.1% acetic acid in water and acetonitrile/ methanol (43:57,v/v) as the mobile phase at a flow rate of 0.6 mL/min. The calibration curve obtained was linear (r2 ≥ 0.9900) over the concentration range of 0.500-500 ng/mL. Method validation was performed as per FDA guidelines and the results met the acceptance criteria. The method was suitable for supporting clinical studies.

Citation

Chen L, Joshi P, Piatkivskyi A, Aguilar K and Lin J. Method Development and Validation for the Determination of Pravastatin in Human Plasma by LC-MS/MS. SM Anal Bioanal Technique. 2017; 2(1): 1008.

Introduction

Pravastatin,hexahydro-6-hydroxy-2-methyl-8-(2-methylbutyryloxy)-1-naphthyl-3,5 dihydroxyheptanoate (Figure 1),

Figure 1: Structure of A) Pravastatin; B) 6-epi-Pravastatin; C) 3α-iso Pravastatin.

a competitive inhibitor of 3-Hydroxy-3-Methylglutaryl Coenzyme A (HMG-CoA) reductase, the rate limiting enzyme in cholesterol biosynthesis and is widely used to treat hypercholesterolemia [1]. Pravastatin is characterized as one of the best, due to the hydroxyl group attached to its decalin ring, which results in a greater hydrophilicity than other HMG-CoA reductase inhibitors [1-3].

Many in vivo and in vitro human and animal studies suggest that active transport mechanisms are involved in the pharmacokinetics of pravastatin. Major metabolites of pravastatin include the 3-alpha-isomer of pravastatin and a glutathione conjugate. Significant human metabolites, 3α-iso pravstatin and 6-epi-pravastatin (Figure 1), may form from a nonenzymatic-acid-catalyzed reaction in the stomach. At least 15 other minor metabolites of pravastatin have been detected [4].

Numerous methods have been developed and validated for detection/estimation of Pravastatin in plasma including high performance liquid chromatography with UV detection, LC/MS/MS [5 14]. In addition to pravastatin, 3α-iso-pravastatin and 6-epi-pravastatin could also be quantified in some assays [15,16]. In this project we attempted to combine previous experience and expertise to create reliable, sensitive, high-throughput LC-MS/MS method to determine pravastatin in human plasma by separating the major isobaric metabolites.

Materials and Methods

Chemicals and reagents

Pravastatin Sodium reference standard was obtained from U.S. Pharmacopeial Convention (Rockville, MD, USA). Pravastatin-D3 sodium salt was used as an Internal Standard (IS) and obtained from Toronto Research Chemicals (Toronto, ON, Canada). Chemical structures are presented in Figure 1. Water used for the LC-MS/MS analysis was produced by in-house ultrapure water system from Millipore (Billerica, MA, USA). Acetonitrile and methanol were of HPLC grade and purchased from Honeywell (Muskegon, MI, USA). Ammoniumformate, formic acid, tert-Butyl methyl ether (MTBE) and acetic acid were analytical grade. The blank human plasma and whole blood were purchased from BioreclamationIVT (Westbury, NY USA).

Preparation of standard solutions and quality control samples

Stock solutions of Pravastatin and Pravastatin-D3 (IS) were prepared in methanol/water (50:50, v/v) at a concentration of 100 µg/mL and stored at 5°C. Working standard solutions of Pravastatin and IS were prepared by dilution of stock solutions in methanol/water (50:50, v/v) and stored at 5°C. The working solutions of Pravastatin were diluted by methanol/water (50:50, v/v) before preparations of 25, 50, 125, 250, 1250, 2500, 12500, 22500 and 25000 ng/mL. The plasma calibration curves were prepared at concentrations of 0.500, 1.00, 2.50, 5.00, 25.0, 50.0, 250, 450 and 500 ng/mL by spiking 490 µL of blank plasma with 10 µL of working solution at appropriate concentration (to make up to 500 µL total sample volume).

The Quality Control (QC) samples were independently prepared in blank plasma at four levels of concentrations for Pravastatin: 0.500 ng/mL (limit of quantification), 1.500 ng/mL (low concentration QC, LQC), 200 ng/mL (medium concentration QC, MQC), and 400 ng/ mL (high concentration QC, HQC).

Sample preparation

The analyte (Pravastatin) and its internal standard (Pravastatin-D3 ) were extracted from human plasma (50 µL) by a liquid liquid extraction procedure. To the matrix blank sample and selectivity samples, 50 µL of methanol/ water (50:50, v/v) was added. To the zero sample, calibration standards, QC samples and analytical samples, 50 µL of the IS solution (100 ng/mL) were added. To the samples, 200 µL of the 0.2 M ammonium formate (pH 3.6) and 3 mL of Tert-Butyl methyl ether were added. The samples were mixed for approximately 5 minute with a vortex mixer after seal followed by centrifugation (e.g. preset value: 2,000×g, 4°C, 5 minutes). The sample tubes were immersed in a dry ice/acetone bath (e.g. for 2 minutes) to freeze the aqueous layer. The organic layer was transferred to glass tubes and evaporated to dryness under a stream of nitrogen gas in a temperature-controlled water bath unit set at 40°C, approx. 15 minutes. The residue was reconstituted in 200 µL of 0.1% acetic acid in water/ [acetonitrile/methanol (43:57, v/v)] (55:45, v/v) by mixing for approximately 30 seconds with a vortex mixer after seal. The reconstituted samples were centrifuged briefly using the “flash” function on an AX-320 centrifuge, and then stored refrigerated (set temperature: 5°C) until analysis.

LC-MS/MS instrument and conditions

Shimadzu Nexera system (SIL-30AC, LC-30AD, DGU-20A5, CTO-20A, Rackchanger II) was used. Mass spectrometric detection was performed on an AB SciexTriple Quad 5500 instrument using MRM. Data processing was performed on Analyst 1.5.1 software package (SCIEX). Detection was performed by Turbo ionspraynegative mode with Unit Resolution. For Pravastatin, the MH- (m/z 423.2) was monitored as the precursor ion and a fragment at m/z 321.2 was chosen as the product ion. For internal standard, the MH- (m/z 426.3) was monitored as the precursor ion and a fragment at m/z 321.1 was monitored as the product ion. Mass parameters were optimized as source temperature 600°C, nebulizer gas (GS1) 40 psi, turbo gas (GS2) 60 psi,curtain gas 30 psi, Collision gas 7 psi, ion spray (IS) voltage -4500 V, declustering potential -110 V, collision energy -20 V, collision cell exit potential -8 V for both analyte and IS

ACE 3 C18 (3 × 50 mm, particle size 3 µm) was selected as the analytical column. The mobile phase composition was 0.1% acetic acid in water and acetonitrile/methanol (43:57, v/v) in the ratio of 55:45 (v/v). The flow rate of the mobile phase was set at 0.6 mL/ min. The column temperature was set at 40°C. The retention time of Pravastatin and IS was found to be 2.4 min approximately at overall 3.5 min run time.

Method validation

A thorough validation of the method was carried out as per the US FDA guidelines [17]. The analytical method was fully validated in human plasma with respect to selectivity, matrix effects, intra- and inter-assay precision and accuracy, recovery, stability in plasma (freeze/thaw, short-term, long-term) post-preparative stability (reinjection reproducibility, extract stability and batch length stability), whole blood stability, dilution integrity, Lower Limit Of Quantitation (LLOQ), carryover, stock and working solution stability.

Selectivity was demonstrated by analysis of blank plasma from a minimum of 6 individual sources (n=1 each) and a blank consisting of a pool (n=1) of the individual sources and a pooled plasma sample (n=1) spiked with concomitant drugs and isomers. In at least 90% of the individual sources and each of the pooled samples, the peak area response of the interfering peaks should be less than 20% of the mean peak area in the acceptable LLOQ calibration standards for the analyte, and less than 5% of the mean peak area of the IS in the acceptable LLOQ calibration standards.

Calibration standards were freshly prepared in duplicate in plasma and both replicates were assayed to generate the calibration curve in at least 3 separate batches (on at least two different days). The linearity (correlation coefficient: r) of the calibration curves and the accuracy of the back-calculated value at each calibration standard level were evaluated. Curves for intra and inter-assay precision and accuracy were used for calibration curve evaluation. The correlation coefficient (r) should not be less than 0.9900. The accuracy for calibration standards should be within ±15% of the nominal value at every concentration except for the LLOQ calibration standards, where it should be within ±20% of the nominal value. At least 75% non-zero standards and a minimum of six calibration standard levels should meet the above criteria in each calibration standard set. Half of LLOQ and ULOQ standards should meet the above criteria for the core intra and inter-assay precision and accuracy.

For intra/inter-assay precision and accuracy, LLQC, LQC, MQC, and HQC (n=6 at each level) were analyzed in three batches on at least two different days. For each batch, the intra/inter-assay precision (CV) and accuracy were calculated for each concentration level, and at least five values must be available for each concentration level to calculate statistics. The CV of each batch should not exceed 15% for each QC level except for the LLQC, where the CV should not exceed 20%. The accuracy of the mean concentration of each batch should be within ±15% of the nominal values for each QC level except for the LLQC, where the mean value should be within ±20% of the nominal value. At least 67% of QC sample accuracy for each batch should be within ±15% of their respective nominal values except LLOQ where the accuracy should be within ±20% of nominal value. No more than 50% of the QC replicates at the same concentration level can be outside of ±15% of the nominal values except LLQC, where the accuracy can be outside of ±20% of nominal value.

Matrix effects were evaluated using three types of plasma; normal plasma, 5% hemolyzed plasma and hyperlipidemia plasma. The plasma samples were analyzed for matrix effects at the LQC and HQC concentration levels (6 sources for each type, n=1 each at each level).

The CV of the IS-normalized Matrix Factor (MF) calculated from the 6 sources of each type of plasma should not exceed 15 % at each concentration level.

The recovery of analyte, Pravastatin, was determined by comparing the peak area ratios (Analyte/IS) of the recovery samples at three concentration levels (LQC, MQC and HQC, n=5 for each level) with the mean peak area ratios of the corresponding reference samples prepared in extracted blank matrix samples (n=5 for each level). The CV of recovery for Pravastatin was calculated for each concentration level and should not exceed 15%. Similarly, the recovery of IS, Pravastatin-D3 , was determined by comparing the corresponding peak area ratios [1/(Analyte/IS)] of the IS recovery samples at HQC level (n=5) with the mean peak area ratio of the corresponding reference samples. The CV of recovery for IS was calculated and should not exceed 15%.

The concentrations of the QC samples (n=3 for LQC and HQC) for stability in plasma were verified immediately after preparation and then again after stability periods. The initial concentration verification can be performed during intra-assay or inter-assay precision and accuracy determination using those samples or during other validation tests. The analyte should be considered stable during a period of storage or throughout a procedure, if the mean concentration of stability samples at each level is within ±15% of the nominal concentration. At least 67% (four out of six) of the batch QC samples must have accuracy within ±15% of their respective nominal values.

Whole blood stability samples ([Low QC sample for whole blood stability, LQCw] and [Mid QC sample for whole blood stability, MQCw], total n=5 at each concentration level) were held under room temperature for 0 and 2 hours. Plasma fraction was separated from the red blood cells and then analyzed. Whole blood stability was evaluated based on the mean peak area ratio of the stored samples. The analyte (Pravastatin) should be considered stable during a period of time or throughout a procedure, if the mean peak area ratio analyte/IS of the stored samples is within ±15% of the mean peak area ratio of 0 hour.

The Dilution QC (DQC) at analyte concentration level of 2000 ng/ mL (n=1) was diluted 10 fold (n=6 each) with blank plasma (pooled),and the diluted samples were analyzed with calibration standards and batch QC samples. The CV and accuracy were calculated based on the measured concentrations and the mean value, respectively. At least f ive values must be available for each level to calculate statistics. The CV should not exceed 15% and the accuracy of the mean value should be within ±15% of the nominal values. At least 67% of DQC sample accuracy should be within ±15% of their respective nominal values.

Results

Method development

Separation of Pravastatin and isomers was attempted using various combinations of acetonitrile, methanol and buffer with varying contents of each component on different columns like ACE, Hypersil, Waters, Zorbax, and Intertsil. It was found that a mixture of 0.1% acetic acid in water and Acetonitrile/methanol (43:57, v/v) with ACE 3 C18 column (3 × 50 mm, particle size 3 µm) could achieve this purpose. The mass spectrum condition was optimized with infusion and Flow Injection Analysis (FIA) of the Pravastatin and IS solutions using Analyst software.

Selectivity

The presence of interfering peaks was determined in the MRM chromatograms for each selectivity sample. In at least 90% of the individual sources and each of the pooled samples, the peak area response of the interfering peaks was less than 20% of the mean peak area in the acceptable LLOQ calibration standards for the analyte, and less than 5% of the mean peak area of the IS in the acceptable LLOQ calibration standards. The results met the validation acceptance criteria.

Selectivity was also demonstrated by analysis of a pooled plasma sample (n=1) spiked with isomers. There was no interference peak at the eluting position of IS, but interfering peak more than 20% of the mean peak area in the acceptable LLOQ calibration standards was observed at the eluting position of Pravastatin. It was confirmed that the eluting positions of 3α-Hydroxy Pravastatin and 6-epi-Pravastatin were well separated from Pravastatin (Figure 2).

Figure 2: MRM chromatogram obtained from sample spiked with Pravastatin Isomers.

The MRM chromatogram for the selectivity sample is presented in Figure 3.

Figure 3: MRM chromatogra obtained from selectivity sample

Calibration curve

The accuracy of at least 75% of calibration standards and a minimum of six calibration standard levels in each calibration standard set were within ±15% (±20% for the LLOQ) of the nominal concentration at every concentration; therefore, calibration standards met acceptance criteria. The linearity (correlation coefficient: r) of each calibration curve met acceptance criteria, as each was >0.9900 (Figure 4 and Table 1).

Figure 4: A Representative Fitted Calibration Curve for Pravastatin in Human Plasma.

Table 1: Calibration Curve Data for Pravastatin in Human Plasma

Nominal Concentration (ng/mL)

Mean observed

concentration (ng/mL)

CV (%)

Accuracy (%)

S1 (0.500)

0.505

5.2

101.0

S2 (1.00)

0.984

6.4

98.4

S3 (2.50)

2.47

3.8

98.7

S4 (5.00)

5.03

2.9

100.5

S5 (25.0)

25.1

1.7

100.3

S5 (50.0)

50.1

2.2

100.3

S6 (250)

247

3.1

98.8

S7 (450)

459

2.4

101.9

S8 (500)

500

2.8

100.1

Representative MRM chromatogram for LLOQ is presented in Figure 5.

 Figure 5: MRM chromatogram obtained from a LLOQ sample (0.500 ng/ mL)

Intra-assay precision and accuracy

LLQC, LQC, MQC, and HQC (n=6 at each level) were evaluated in three batches on at least two different days. LLQC (n=6) was analyzed in one batch. The intra-assay data met acceptance criteria, with %CV ≤15% for each QC level (≤20% for LLQC) and accuracy within ±15% of the nominal value for each QC level (±20% for LLQC) and at least 67% of QC sample accuracy for each batch should be within ±15% of their respective nominal values (±20% for LLQC). The intra-assay precision and accuracy results are presented in table 2.

Table 2: Intra/Inter-assay Precision and Accuracy for Pravastatin in Human Plasma.

 

QC

Sample

Mean observed

concentration

CV

Accuracy

(ng/mL)

(ng/mL) (n=6)

(%)

(%)

 

 

Intra-assay precision and accuracy

LLQC

(0.500)

0.518

2.7

103.7

LQC (1.50)

1.54

1.5

102.5

MQC (200)

200

4.2

99.8

HQC (400)

391

2.4

97.8

 

 

Inter-assay precision and accuracy

LLQC

(0.500)

0.52

5.1

104.0

LQC (1.50)

1.51

3.5

100.6

MQC (200)

200

3.1

99.8

HQC (400)

393

1.9

98.2

Inter-assay precision and accuracy

Inter-assay accuracy and precision were evaluated from the plasma quality control pool (LLQC, LQC, MQC and HQC) in 3 validation runs over three separate batches on at least two different days, where each QC was analyzed n=6 per run. The intra-assay precision and accuracy results for human plasma were also used as the inter-assay precision and accuracy results. The inter-assay precision and accuracy data met the acceptance criteria. The overall %CV was <15% for each QC level (≤20% for LLQC). The accuracy of the overall mean concentration was within ±15% of the nominal values for each QC level (±20% for LLQC). More than 67% of QC sample’s accuracy for each batch at each level was within ±15% of their respective nominal values (±20% for LLQC). The inter-assay precision and accuracy results are presented in table 2.

Matrix effects

Matrix effects were evaluated by analysis of 6 sources ([3 males and 3 females], n=1 each at each concentration level) of blank matrix (normal plasma, 5% hemolyzed plasma and lipemic plasma) at the LQC and HQC concentration levels. IS normalized MF was calculated for each analyte and source of matrix, this was done by calculating the ratio of the peak area ratio in the presence of matrix (measured by analyzing blank matrix spiked with analyte after extraction), to the peak area ratio in absence of matrix (pure solution of the analyte). The %CV of the IS-normalized MF calculated from the 6 sources of matrix did not exceed 15 %. The results met the validation acceptance criteria.

Recovery

The recovery of analyte was determined by comparing the peak area ratios (analyte/IS) of the recovery samples at three concentration levels (LQC, MQC and HQC, n=5 for each level) with the mean peak area ratios of the corresponding reference samples (n=5 for each level). Similarly, the recovery of IS was determined by comparing the peak area ratios [1/(analyte/IS)] of the IS recovery samples at HQC level (n=5) with the mean peak area ratio of the corresponding reference samples. The CV for each concentration level did not exceed 15% and the results met the acceptance criteria.

Freeze/Thaw stability

The influence of the physical processes of freezing and thawing on analyte stability in biological matrix was evaluated by analyzing plasma LQC and HQC stability samples (n=3 for each level) that were subjected to 5 freeze/thaw cycles prior to analysis. A freeze/thaw cycle consists of keeping the quality control samples frozen for at least twelve hours at -80°C and at -20°C and then completely thawing at room temperature (approximately 25°C), with the initial freezing period lasting at least 24 hours. The freeze/thaw stability data met the validation acceptance criteria, with the accuracy of the mean value within ±15% of the nominal value for each quality control level. The results of freeze/thaw stability are presented in Table 3.

Table 3: Stability for Pravastatin in Human Plasma.

 

Nominal concentration

(ng/mL)

Mean observed

concentration (ng/mL) (n=3)

Accuracy (%)

Room temperature stability in plasma (24 h)

LQC (1.50)

1.48

98.6

HQC (400)

400

99.9

 

Extraction stability (72 h)

LQC (1.50)

1.62

108.1

HQC (400)

399

99.6

 

Re-injection stability (72 h)

LQC (1.50)

1.51

100.4

HQC (400)

392

98.0

 

Freeze/thaw stability in plasma (5 cycles, -20°C and

-80°C)

LQC (1.50)

1.62

108.3

HQC (400)

393

98.3

LQC (1.50)

1.53

101.8

HQC (400)

387

96.8

 

 

Long term stability in plasma (178 days, -20°C and -80°C)

LQC (1.50)

1.47

98.0

HQC (400)

400

99.9

LQC (1.50)

1.52

101.5

HQC (400)

393

98.3

The results indicate that the analyte is stable in matrix for 5 freeze and thaw cycles when stored at either -20°C or -80°C.

Short-term stability

Short-term stability was evaluated by analyzing plasma stability samples at LQC and HQC levels (n=3 for each level) that were stored under frozen conditions (-80°C) for at least 24 hours, completely thawed at room temperature (approximately 25°C), and then subsequently stored at room temperature for approximately 24 hours. The short-term stability data met the validation acceptance criteria, with the accuracy of the mean value within ±15% of the nominal value for each quality control level. The results of short term stability are presented in Table 3. The results indicate that the analyte is stable in matrix stored at room temperature till 24 hours.

Long-term Stability

Long-term stability was evaluated by analyzing plasma stability samples at LQC and HQC levels (n=3 for each level) that were stored under frozen conditions (-20°C and -80°C) for 6 months and then processed and analyzed for long-term stability. The long-term stability data met the validation acceptance criteria, with the accuracy of the mean value within ±15% of the nominal value for each quality control level. The results of long-term stability are presented in Table 3. The results indicate that the analyte is stable in matrix when stored in a freezer set at -20°C (acceptable temperature: -25°C to -15°C), and in another freezer set at -80°C (acceptable temperature: -90°C to -70°C) till 178 days

Batch length stability

Batch length stability was demonstrated by analyzing a run containing MB, Z, calibration standards, and LQC, MQC and HQC samples (n=1 at each level), with the number of expected samples represented by injecting MB samples or additional QC replicates, and a second set of QC samples, MB, Z and calibration standards. A series of MB samples and additional QC replicates were injected to mimic the anticipated maximum number of unknown samples to be analyzed in one sample analysis batch. The batch length stability results indicate that maximum 120 samples can be analyzed in one run without any effect on stability

Re-injection stability

Re-injection stability was demonstrated by re-injecting the previously injected LQC and HQC intra-assay reproducibility samples (n=6 for each level), as well as the corresponding previously injected calibration standards, after storage in the sample compartment refrigerator (set at 5°C) for at least 72 hours after the initial injection. The results met the validation acceptance criteria for all calibration standard levels. The re-injection reproducibility results met the validation acceptance criteria, with %CV ≤15% and the accuracy of the mean value within ±15% of the nominal value for each QC level. T he results of re-injection reproducibility are presented in Table 3. These results indicate that the analyte in processed and previously injected samples is stable for at least 72 hours in an auto sampler set at 5°C.

Extract stability

Extract stability was demonstrated by extracting LQC and HQCstability samples (n=3 for each level) and analyzing them after being stored in the auto sampler (set at 5°C) at least 72 hours and analyzed with freshly prepared calibration standards. The extract stability results met the validation acceptance criteria, with %CV ≤15% and the accuracy of the mean value within ±15% of the nominal value for each QC level. The results of extract stability are presented in Table 3. Theresults indicate that the analyte in processed samples is stable till 72 hours in an autosampler set at 5°C.

Stock and standard solution stability

Stability under room temperature and frozen conditions for stock and standard solutions were evaluated, and were based on the mean peak area ratios and mean IS peak area ratios [1/(analytePravastatin/ IS)] obtained.Pravastatin and IS stock solutions and working solutions were held at room temperature (approximately 25°C) for approximately 24 hours, and then injection samples (n=3) for stock solution and working solutions stability were prepared from the stored solutions. Injection samples (n=3) for stock solution and working solutions stability were prepared from freshly prepared solutions (for initial values). The stability of stock solutions and working solutions met the acceptance criteria, stability mean values were within ±15% of the initial mean value and the results are presented in Table 4.

Table 4: Stock Solution and Working Solution Stability of Pravastatin.

 

Nominal concentration

Period

Remaining (%) (n=3)

Stock solutionroom temperature

stability

100 µg/mL

24 hours

100.7

Stock solution stability at 5°C

100 µg/mL

184

days

100.4

IS Stock solutionroom temperature

stability

100 µg/mL

24 hours

95.8

IS Stock solution stability at 5°C

100 µg/mL

184

days

89.6

Working solutionroom temperature

stability

25 ng/mL

24 hours

98.4

Working solution stability at 5°C

25 ng/mL

184

days

100.6

IS solutionroom temperature

stability

100 ng/mL

24 hours

97.8

IS solution stability at 5°C

100 ng/mL

184

days

93.7

The results indicate that the stock solution and working solutions for Pravastatin and IS can be stored at room temperature for at least 24 hours.

Pravastatin and IS stock solutions and working solutions were held under frozen conditions (acceptable temperature: 2°C to 8°C) for approximately 6 months, and then injection samples (n=3) for stock solution and working solutions stability were prepared from the stored solutions. Injection samples (n=3) for stock solution stability and working solutions were prepared from freshly prepared solutions (for initial values). The stability of stock solutions and working solutions met the acceptance criteria, stability mean values were within ±15% of the initial mean value and the results are presented in Table 4. The results indicatethat the stock solution and working solutionsfor Pravastatin and IS can be stored at refrigerated condition for at least 184 days.

Whole blood stability

The 2 hour whole blood stability samples met the acceptance criteria for Pravastatin, stability mean values were within ±15% of the mean peak area ratio of the 0 hour sample. The results indicate that the analyte is stable in whole blood for at least 2 hours for Pravastatin.

Dilution integrity

The dilution integrity evaluation met the acceptance criteria, with %CV ≤15% for replicates and the accuracy within ±15% of the nominal value for the mean value. The results indicate that plasma samples can be diluted with plasma without affecting precision and accuracy.

Determination of LLOQ

The Lower Limit of Quantitation (LLOQ) is the lowest non-zero human plasma quality control pool concentration level which can be accurately and reproducibly quantified. The LLQC was analyzed (n=6) and the data met acceptance criteria, with %CV ≤20% and the mean accuracy was within ±20% of the nominal value and at least 67% of the QC samples were within ±20% of the nominal value. In addition, the analyte peak area response at the LLOQ was at least five times the response compared to the first MB peak response and the signal to noise ratio was at least 5. For this validation, the LLOQ was determined to be 0.500 ng/mL.

Carryover

The potential for analyte carryover from high-level to low-level samples in an injection sequence was evaluated by injecting duplicate extracted matrix blanks immediately after the ULOQ calibration standards in an intra-assay precision and accuracy validation run. Carryover exceeding acceptance criteria was not observed, as the carryover peak area was less than 20% of the peak area of Pravastatin at LLOQ and no greater than 5% of the peak area of IS on the MRM chromatograms in the first matrix blanks injected immediately after the ULOQ calibration standards. The carryover results met acceptance criteria.

Batch acceptance

Stability analysis in matrix (except for whole blood stability and re-injection stability) and Dilution integrity were performed with freshly prepared calibration curves and freshly prepared batch QC samples (n=2 at LQC, MQC and HQC) unless intra- or inter-assay precision and accuracy QC samples were used as batch QC samples. At least 67% of accuracy for QC samples (n=2 each, total 6) for stability evaluation run were within ±15% of their respective nominal values. No more than 50% of QC samples for batch acceptance at the same concentration were outside ±15% of the nominal values.

Discussion

 In the present investigation, we have developed a method with liquid-liquid extraction technique. The following are the advantages of the proposed method over those reported earlier: (1) Because of the use of less plasma volume (50 µL), the volume of the sample to be collected per time point from an individual during the study is reduced significantly; (2) Greater sensitivity is achieved even with low plasma volumes and the method is well suited for pharmacokinetic analysis; (3) Separation of Pravastatin from isobaric metabolites excluded the inference from the plasma samples.

Conclusion

In summary, the method is described for the quantification of pravastatin from human plasma by LC-MS/MS in negative ionization mode using multiple reaction monitoring. The current method has shown acceptable precision and adequate sensitivity for the quantification of pravastatin in human plasma samples obtained for pharmacokinetic studies. The method described is simple, rapid, sensitive, selective and fully validated according to commonly accepted criteria. The cost-effectiveness and simplicity of the assay make it an attractive procedure in high-throughput bioanalysis of pravastatin.

References

1.Hatanaka T. Clinical pharmacokinetics of pravastatin: mechanisms of pharmacokinetic events. Clin Pharmacokinet. 2000; 39: 397-412

2. Lennernäs H, Fager G. Pharmacodynamics and pharmacokinetics of the HMG-CoA reductase inhibitors. Similarities and differences. ClinPharmacokinet. 1997; 32: 403-425.

3. Haria M, McTavish D. Pravastatin. A reappraisal of its pharmacological properties and clinical effectiveness in the management of coronary heart disease.Drugs. 1997; 53: 299-336.

4. Wehr, A. Pravastatin. Handbook of Metabolic Pathways of Xenobiotics. 2014; 2: 1-6.

5. Otter K, Mignat C. Determination of pravastatin in human plasma by high performance liquid chromatography with ultraviolet detection. J Chromatogr B Biomed Sci Appl. 1998; 708: 235-41

6. Bauer S,Mwinyia J, Stoecklea A, Gerloffa T, Roots I. Quantification of pravastatin in human plasma and urine after solid phase extraction using high performance liquid chromatography with ultraviolet detection. J Chromatogr B AnalytTechnol Biomed Life Sci.2005; 818 : 257-262.

7. Önal A,Sagirli O. Development of a Selective LC Method for the Determination of Pravastatin Sodium.Chromatogrhaphia. 2006; 64: 3-4.

8. Sparidans RW,Iusuf D, Schinkel AH, Schellens JH, Beijnen JH. Liquid chromatography-tandem mass spectrometric assay for pravastatin and two isomeric metabolites in mouse plasma and tissue homogenates. J Chromatogr B AnalytTechnol Biomed Life sci.2010; 878: 2751-2759.

9. Tan ZR,Ouyang DS, Zhou G, Guo D, Chen Y, et al. A simple and sensitive LC-MS/MS method for determination of pravastatin in human plasma. Chinese J New Drugs. 2008; 17: 1150-1153

10. Deng JW, Kim KB, Song IS, Shon JH, Zhou HH, et al. Determination of two HMG-CoA reductase inhibitors, pravastatin and pitavastatin, in plasma samples using liquid chromatography-tandem mass spectrometry for pharmaceutical study.Biomed Chromatogr. 2008; 22: 131-135. Citation: Chen L, Joshi P, Piatkivskyi A, Aguilar K and Lin J. Method Development and Validation for the Determination of Pravastatin in Human Plasma by LC-MS/MS. SM Anal Bioanal Technique. 2017; 2(1): 1008. Copyright  Chen L

11. Jain DS,Subbaiah G, Sanyal M, Jain VK, Shrivastav P. A rapid and specific approach for direct measurement of pravastatin concentration in plasma by LC-MS/MS employing solid-phase extraction.Biomed Chromatogr.2007; 21: 67-78.

12. Kyrklund C,Backman JT, Neuvonen M, Neuvonen PJ. Effect of rifampicin on pravastatin pharmacokinetics in healthy subjects.Br J Clin Pharmacology.2004; 57: 181-187.

13. Kawabata K,Samata N, Urasaki Y.Quantitative determination of pravastatin and R-416, its main metabolite in human plasma, by liquid chromatography tandem mass spectrometry. J Chromatogr B AnalytTechnol Biomed Life Sci. 2005; 816: 73-79.

14. Mertensa B,Cahaya B, Klinkenberga R, Streel B.An automated method for the simultaneous determination of pravastatin, 3-hydroxy isomeric metabolite, pravalactone and fenofibric acid in human plasma by sensitive liquid chromatography combined with diode array and tandem mass spectrometry detection. J Chromatogr A. 2008; 1189: 493-502.

15. Zhang X, Vernikovskaya DI, Wang X, Nanovskaya TN, Costantine M, Hankins GD, Ahmed MS.Quantitative determination of pravastatin and its metabolite 3α-hydroxy pravastatin in plasma and urine of pregnant patients by LC-MS/ MS.BiomedChromatogr. 2016; 30: 548-54

16. Mulvana D, Jemal M, PulverSC.Quantitative determination of pravastatin and its biotransformation products in human serum by turbo ion spray LC/MS/ MS.J Pharm Biomed Anal. 2000; 23: 851-66.

17. US DHHS, FDA, CDER. Guidance for Industry: Bioanalytical Method Validation. U.S. Department of Health and Human Services, Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CV). 2001

Other Articles

Article Image 1

Drug Optimization: Fighting Research to Achieve the Greatest Use of Darunavir

A design has been development that provides higher activity to darunavir, an antiretroviral drug, aiming a lower dose in the battle against HIV in children and adults. Globally this research paves the way to get a new darunavir: β-cyclodextrin complex driving future perspectives to new anti HIV drugs for clinical applications. β-cyclodextrins not only promote the solubility of darunavir, but can drive to development of new complexes able to combat HIV in lower doses and, therefore lower toxic effects. Analysis methods by chromatography, also, were developed to evaluated the quality of the complexed darunavir. All this effort to contemplate the patients of all ages and provide quality medicines as well as a better quality of life.

Ana Carolina Kogawa¹ and Hérida Regina Nunes Salgado¹*


Article Image 1

Enhancing Throughput of Glutathione Adduct Formation Studies and Structural Identification Using a Software-Assisted Workflow Based on High-Resolution Mass Spectrometry (HRMS) Data

The bioactivation of drugs to Reactive Metabolites (RM) has been related to drug-induced liver injury and hypersensitivity reactions in patients. Therefore, many pharmaceutical companies are investigating the potential to form reactive metabolites in vitro as an integral part of the optimization of drug candidates. A computer assisted workflow to efficiently analyze larger numbers of compounds for the formation of glutathione trappable RM is presented here. A set of 95 compounds with known bioactivation potential was selected for this study. Incubations with human liver microsomes were prepared with GSH. The acquisition of MS/MS spectra was triggered by ion intensity. MS with singly and doubly charged ions were used for peak detection and MS/MS spectra were used for structural elucidation. A confidence classification system for the GSH peak detection (high, medium, low) was developed based on the detection of characteristic fragment ions or neutral losses and applied to remove potential false positive results. A comparative analysis of the HRMS results with literature data was carried out. The most frequently observed Neutral Loss (NL) found in singly charged GSH adducts (drug-glutathione conjugates) were, the Neutral Loss (NL, 129 Da) and Fragment Ion (FI, m/z 308) and in the doubly charged ones the Fragment Ion (FI, m/z 130). These NL and FI were used to identify GSH-related drug metabolites. MS/MS spectra were inspected to aid structural elucidations: 17% of drug substrates and 29 % of GSH adduct metabolites were identified with only doubly charged ions, stressing the importance of considering this charge state in the identification workflow. A total of 41 compounds that form GSH adducts were retrieved from literature (HRMS, identified 28 compounds (68%) in high confidence, and the same result was obtained using precursor ion scan). By the confidence analysis of GSH peaks, the quality of the each GSH adduct was determined.

Esra Nurten Cece-Esencan¹, Fabien Fontaine², Guillem Plasencia²,³, Marieke Teppner³, Andreas Brink³, Axel Pähler³, and Ismael Zamora⁴*


Article Image 1

Evaluation of Solid Phase Extraction NH2 for Determination of Enniatins (A, A1, B, B1) and Beauvericin in Cereals by UPLC-MS/MS

We report herein, for the first time, the application of NH2 solid-phase extraction NH2 (NH2-SPE) cartridge for cleanup of enniatins (ENA, ENA1 , ENB, ENB1 ) and Beauvericin (BEA) in cereals. Samples were sequentially extracted with a solution containing 80% acetonitrile and 0.1% formic acid and the extracts were cleaned up with NH2-SPE cartridges before being analyzed by Ultra-High Performance Liquid Chromatography-Mass Spectrometry (UPLC-MS/MS). A CORTECS C18 column was used for separation of enniatins and beauvericin. Mass spectrometric analysis was conducted at the Positive Electrospray Ionization (ESI+) Mode with Multi Reactions Monitoring (MRM). Very good linear relationships between spiked levels of ENA, ENA1, ENB, ENB1, BEA in cereals and mass spectra were observed with regression coefficients of 0.995 to 0.999 and Detection Limits (LOD) range of 0.01 to 0.1 µg kg-1. Clean up with NH2-SPE cartridge gave more satisfactory recoveries ranging from 80.9% to 109.8% than with QuEChERS method that gave recoveries between 63.5% to 96.6%. Application of the NH2-SPE cartridge for the determination of enniatins and beauvericin showed detection levels ranging from 0.01 µg kg-1 to 127.87 µg kg-1 in corn and 0.01 µg kg-1 to 116.59 µg kg-1 for wheat.

Juan Sun, Weixi Li, Yan Zhang, Huijie Zhang, Li Wu, Xuexu Hu and Bujun Wang*


Article Image 1

Electroanalysis: Towards a

Recent trends in the sensor community, indicates that one of the major challenges and opportunities of the field relies on developing smart sensor platforms, which are cheap, efficient, easy-to-use, and capable of minimizing tasks at the end user stage, in comparison to traditionally used methods which require bulky instrumentations and qualified personnel.

Stefano Cinti*


Article Image 1

Chemistry in Sustainability and Chemistry of Sustainability: Waste of use of Fishing Industry for Removal of Waste Textile Industry

One of the negative consequences related to socioeconomic development in recent centuries has been the rise of environmental contamination. In light of this, human has sought to implement actions in all spheres (social, economic and cultural) aiming to minimize the impact on the environment. Concern about the pollution of bodies of water and industrial wastewater treatment is the biggest worry in this regard. Among various treatment methods highlights the adsorption solid phase, by being very efficient and economically feasible. The use of chitosan as an adsorbent makes the process more ecologically interesting, once chitosan is also retrieved from waste. In this study we evaluate the adsorption process of the bromothymol blue dye by the chitosan. The parameters that influence the sorption process were studied (pH, ionic strength, adsorbent mass and dye initial concentration), and the process thermodynamics.

Francisco JE¹, Cajé JCM¹, Semaan FS¹, and Pacheco WF¹*


Article Image 1

Sorption Properties of Methyl Orange onto Chemically Modified Chitosan: Thermodynamics and Kinetics Studies

A new kind of sorbent with high adsorption capacity and stability was prepared using chitosan and epichlorohydrin through a cross-linking reaction and assessed by using methyl orange. The sorption capability was evaluated by means of kinetics and equilibrium studies. Relevant factors for such sorption process were also assessed. After optimization of the relevant sorption factors, the percentages for methyl orange removal by chitosan and by modified chitosan were 82%, and 67%, respectively. Although the cross-linked chitosan offered the lowest removal percentage, this sorbent presented other relevant advantages when compared to non-modified chitosan, such as higher chemical and thermal stabilities.

Juliana CM Caje, Paula Marcelle de Oliveira, Felipe S Semaan, Raphael C Cruz, Ricardo J. Cassella and Wagner F. Pacheco


Article Image 1

Quantitative Headspace Solid-Phase Microextraction Gas Chromatography Mass Spectrometry (HS-SPME-GC-MS/ MS) Method for the Determination of Tributyltin in Sediment: Validation according to EU Directive Requirement

The Directive 2013/39/EU has reviewed the list of priority substances under the European Water Framework Directive (WFD; Directive 2000/60/EC) and has strengthened the principle of flexibility for Member States in applying the Environmental Quality Standard (EQS) for alternative matrices. Coherently, it has established new EQS for biota and it has invited Member States to set EQS for more opportune matrices, able to advantage the monitoring strategy and offer the same level of protection, such as sediment. The present work developed a (HS)SPME-GC-MS/MS method for the analysis of Tributyltin (TBT) in sediment samples and, through an in depth validation process, it assessed the compliance with the technical specification required by the EU Directive 2009/90/EC for chemical status analysis under the scope of the WFD. The EQS established for TBT in sediment by the Italian Environmental Ministry was used as reference. The method fulfils the minimum performance criteria required by the EU Directive 2009/90/EC (Method Detection Limit (MDL) and Minimum Level Of Quantitation (ML), expressed as ng Sn g-1 d.w., were 0.2 and 0.5, respectively). More generally, the analytical figures of merit achieved, were satisfying for the target TBT concentration range (recovery: 90-111%, intermediate precision range: 6-12%).

Seta Noventa¹, Malgorzata Formalewicz¹, Jvan Barbaro², Claudia Gion¹, Federico Rampazzo¹, Massimo Gabellini³, Rossella Boscolo¹, and Daniela Berto¹*

 


Article Image 1

Cramer Rao Lower Bound CRB Shape Detection Method

In this paper we investigate the problem of tracking a moving but non-vibrating cylindrical object and estimating its size and shape using an artificial lateral line system, an d determine the lower bound that the estimator can be reached. Based on a nonlinear analytical model for the moving object-induced flow field, a two-stage extended Kalman filter is proposed to estimate the location, velocity, size, and shapes of the object. Simulation results on tracking an ellipsoidal cross-section are presented to illustrate the approach.

Ahmad T Abdulsadda*


Article Image 1

Composition and Antioxidant Activity of Vitex agnus-castus L and Rosmarinus Officinalis L. Leaves Essential Oils Cultivated in Syria

Medicinal plants contain a wide variety of chemicals which have very important roles in numerous applications including medicinal and those related with industry. Essential oils represent valuable sources for natural antioxidants. The aim of our study was to evaluate the chemical composition and antioxidant activity of the essential oils extracted from leaves of Syrian Vitex agnus-castus L. and Rosmarinus officinalis L., where is both essential oils were extracted and analyzed by gas chromatography-mass spectrometry. The antioxidant activities of these essential oils were determined by three different test systems, scavenging effect on 2,2-diphenyl-1-picrylhydrazyl radical, total phenolic and flavonoids contents. The main constituents found in Vitex agnus-castus L. essential oil were 1,8-Cineole (19.34%) and Sabinene (12.50%), while the major constituents in Rosmarinus officinalis L. essential oil were 1,8-Cineole (28.03%) and α-Pinene (14.70%). The results showed that 2,2-diphenyl-1-picrylhydrazyl radical scavenging and total phenolic contents of Vitex agnus-castus L. essential oil were higher than Rosmarinus officinalis L. essential oil. Total flavonoids contents were not detected in both essential oils. The Vitex agnus-castus L. and Rosmarinus officinalis L. essential oils are sources of natural antioxidants. Therefore, further work is needed to identify the compound(s) responsible for the antioxidant activity of Vitex agnus-castus L. and Rosmarinus officinalis L. essential oils.

Faten Al Saka, Manal Daghestani, and Francois Karabet*


Article Image 1

A Study on Semi-Late Maize Hybrids for Resistance to Smut of Maize /Ustilago maydis/

The most efficient method of controlling diseases of economic importance in agricultural crops is breeding of resistant varieties. The objective of the study was to determine the extent of the attack of smut of maize /Ustilago maydis/ on 22 hybrids and 4 maize standards under conditions, conducive to the spread of the pathogen.

The study was conducted in the experimental fields of the Institute of Agriculture and Seed Science “Obraztsov Chiflik”, Ruse during the period 2012 - 2014. 26 germplasms of Preliminary Variety Trial N 2 (PVT) were included.

Ustilago maydis is a common disease in most parts of the world, where maize is grown. In all above-ground organs of the plant, tumors of smutty mass are formed. The pathogen attack leads to stunted growth of the host plant, reducing yield, and economic losses.

The analysis of data about the new hybrids of IASS “Obraztsov Chiflik” - Ruse showed that fifteen of them were resistant.

The resistance was the highest in the hybrids: St.20, Eks.12, Eks.10, Eks.23, Eks.16, Eks.2, Eks.3, Eks.4, Eks.6, Eks.9, Eks.13, Eks.17, Eks.19, Eks.20, 2 Eks.6. They may be used as sources of resistance. Susceptible were: Eks.15, Eks.11, Eks.8, Eks.24, Eks.1, Eks.5, St.7, Eks.14, Eks.18, Eks.22, Eks.25.

Iliana Ivanova, Lyubomir Ivanov, and Gergana Ivanova*