Validation of Assay Indicating Method Development of Imatinib in Bulk and Its Capsule Dosage Form by Liquid Chromatography

[ ISSN : 2473-0696 ]

Abstract Citation Introduction Experimental Method validation Results and Discussion Conclusion References
Details

Received: 23-Sep-2015

Accepted: 10-Oct-2015

Published: 20-Oct-2015

Nalini Kanta Sahoo¹, Madhusmita Sahu¹, V Alagarsamy¹, B Srividya², and Chinmaya Keshari Sahoo³*

 

 

¹ MNR College of Pharmacy, Osmania University, India

² Yalamarty Pharmacy College, Andhra University, India
³ Osmania University College of Technology, Osmania University, India

Corresponding Author:

Nalini Kanta Sahoo, MNR College of Pharmacy, Osmania University, India, Email: sahoo.nalini@gmail.com

Keywords

RP-HPLC; Imatinib; Quality control level

Abstract

A novel, simple and economic reverse phase High Performance Liquid Chromatography (RP-HPLC) method has been developed for the quantification of Imatinib in bulk and capsule dosage form with greater precision and accuracy. Separation was achieved on Analytical technologies, C-18, (250mm*4.6mm) column in isocratic mode with mobile phase consisting of acetonitrile: potassium dihydrogen phosphate buffer (pH 2.5) (30:70v/v) with a f low rate of 0.8 mL/min. The detection was carried out at 268 nm. The retention time of Imatinib was found to be 2.67 min. The method was validated as per ICH guidelines. Linearity was established for Imatinib in the range 5-35 μg / ml with r2 value 0.996. The percentage recovery of Imatinib was found to be in the range 99.49-99.67 %. The high recovery and low relative standard deviation confirm the suitability of the proposed method for the estimation of the drug in bulk and capsule dosage forms. Validation studies demonstrated that the proposed RP-HPLC method is simple, specific, rapid, reliable and reproducible for the determination of Imatinib for quality control level.

Citation

Sahoo NK, Sahu M, Alagarsamy V, Srividya B and Sahoo CK.Current Status of Two-Dimensional Gel Validation of Assay Indicating Method Development of Imatinib in Bulk and Its Capsule Dosage Form by Liquid Chromatography. Ann Chromatogr Sep Tech. 2015;1(2):1010.

Introduction

Imatinib is chemically a N-(4-methyl-3-{[4-(pyridin-3-yl) pyrimidin-2-yl] amino}phenyl)-4 [(4-methylpiperazin-1-yl) methyl] benzamide. It is an anti neoplastic agent used to treat chronic myelogenous leukemia. Imatinib is used to treat Chronic Myelogenous Leukemia (CML), Gastrointestinal Stromal Tumors (GISTs) and a number of other malignancies. imatinib mesylate was effective in patients with systemic mastocytosis, including those who had the D816V mutation in c-Kit.

However, since imatinib binds to tyrosine kinases when they are in the inactive configuration and the D816V mutant of c-Kit is constitutively active, imatinib does not inhibit the kinase activity of the D816V mutant of c-Kit. Imatinib is a 2-phenylaminopyrimidine derivative that functions as a specific inhibitor of a number of tyrosine kinase enzymes. In chronic myelogenous leukemia, the Philadelphia chromosome leads to a fusion protein of Abl with Bcr (breakpoint cluster region), termed Bcr-Abl. As this is now a continuously active tyrosine kinase, Imatinib is used to decrease Bcr-Abl activity.

Imatinib mesylate is a protein-tyrosine kinase inhibitor that inhibits the Bcr-Abl tyrosine kinase, the constitutive abnormal tyrosine kinase created by the Philadelphia chromosome abnormality in Chronic Myeloid Leukemia (CML). It inhibits proliferation and induces apoptosis in Bcr-Abl positive cell lines as well as fresh leukemic cells from Philadelphia chromosome positive chronic myeloid leukemia.Imatinib also inhibits the receptor tyrosine kinases for Platelet Derived Growth Factor (PDGF) and Stem Cell Factor (SCF)-called c-kit.Imatinib was identified in the late 1990s by Dr Brian J.Druker. Its development is an excellent example of rational drug design.Soon after identification of the bcr-abl target, the search for an inhibitor began. Chemists used a high throughput screen of chemical libraries to identify the molecule 2-phenylaminopyrimidine. This lead compound was then tested and modified by the introduction of methyl and benzamide groups to give it enhanced binding properties, resulting in imatinib [1].

Several analytical methods for the determination of imatinib by spectrophotometry [2,3], HPLC [4-12] and LC/MS [13] have been reported. The aim of the present work was to develop and validate a better sensitive RP-HPLC method that can be implemented for the quantification of imatinib in bulk as well as in its tablet dosage forms when compared to the data of previous established method [4-12].

Experimental

Materials

Imatinib Mesylate working standard was purchased from celon labs,Hyderabad,India.Imatinib Mesylate capsules containing 20 mg of Imatinib (Omef) were obtained from Apollo Pharmaceuticals Pvt. Ltd, Visakhapatnam. Acetonitrile (HPLC grade) were purchased from Qualigens, Potassium dihydrogen orthophosphate, Orthophosphoric acid are purchased from Sd fine-Chem ltd; Mumbai.

Instrumentation

Analytical technologies Alliance High Pressure Liquid Chromatograph installed with Empower software Model: e2487 dual absorbance, phenomenex, C-18, (250mm*4.6mm) column with UV – Vis spectophotometer model UV-2450 were used.

Chromatographic conditions

The High Performance Liquid Chromatographic (HPLC) system used was operated with the column temperature maintained at 300 C,using a mobile phase composition of acetonitrile and Potassium dihydrogen phosphate buffer (pH adjusted to 2.5 with O-Phosphoric acid) in the ratio of 30:70 v/v at a flow rate of 0.8 mL/min within a run time of 10 min.Prior to use, the mobile phase was degassed by an ultrasonic bath and filtered by a Millipore vacuum filter system equipped with a 0.45 μm high vacuum filter.The drug was detected and quantified at 268 nm.

Preparation of standard solutions

25mg of Imatinib Mesylate Working standard was accurately weighed and transferred into a 25 mL volumetric flask and about 20 ml of diluent was added to it and sonicated to dissolve drug completely and volume was made up to the mark with the same solvent which gave Stock solution of 1000 ppm.1 ml of the above stock solution was pippetted into a 10ml volumetric flask and was diluted up to the mark with diluents to prepare 100ppm solution. Further 1 ml of prepared 100 ppm solution was pippetted into a 10ml volumetric flask and was diluted up to the mark with diluents which gave 10ppm Imatinib Mesylate working standard solution. The solution was mixed well and f iltered through 0.45µm filter.

Assay of Imatinib from Marketed capsules

Twenty capsules (Omef capsules, Label claim: Each film-coated capsule contains: Imatinib 20 mg) were taken and the I.P. method was followed to determine the average weight. Above weighed capsules were finally powdered and triturated well. A quantity of powder equivalent to 82.5 mg of drugs were transferred to 100 ml volumetric f lask, and 70 ml of HPLC grade methanol was added and solution was sonicated for 15 minutes, there after volume was made up to 100 ml with same solvent. Then 10 ml of the above solution was diluted to 100 ml with HPLC grade methanol.The solution was filtered through a membrane filter (0.45 mm) and sonicated to degas. From this stock solution (3.5 ml) was transferred to five different 10 ml volumetric f lasks and volume was made up to 10 ml with same solvent system.The solution prepared was injected in five replicates into the HPLC system and the observations were recorded. A duplicate injection of the standard solution was also injected into the HPLC system and the peak areas were recorded.

Method validation

The method was validated in accordance with ICH guidelines [14].The parameters assessed were linearity, accuracy, and Limit of Detection (LOD), limit of quantification (LOQ), precision, reproducibility, robustness and system suitability.

Accuracy

Accuracy was best determined by the standard addition method. Previously analyzed samples of Imanitib API were added with standard drug solutions and are analyzed by the proposed method. Recovery (%), RSD (%) and bias (%) were calculated for each concentration.

Precision

Precision was determined as both repeatability and intermediate precision, in accordance with ICH guidelines. Repeatability of sample injection was determined as intraday variation and intermediate variation. For these determinations, single concentration (10 µg/ ml) at different time intervals and different days, of the solution of Imatinib API was used. The precision of each method was ascertained separately from the peak areas & retention times obtained by actual determination of five replicates of a fixed amount of drug. Imatinib (API) The percent relative standard deviation was calculated for Imatinib.

Robustness

The concept of robustness of an analytical procedure has been defined by the ICH as “a measure of its capacity to remain unaffected by small but deliberate variations in method parameters”.To determine the robustness of the method experimental conditions are purposely altered and chromatographic characters are evaluated. Influence of small changes in chromatographic conditions such as change in flow rate, wavelength of detection and acetonitrile content in mobile phase were studied to determine the robustness of the method.

Limit of Detection (LOD)

The Limit of Detection (LOD) of an analytical method may be defined as the concentration, which gives rise to an instrument signal that is significantly different from the blank. For spectroscopic techniques or other methods that rely upon a calibration curve for quantitative measurements, the IUPAC approach employs the standard deviation of the intercept (Sa), which may be related to LOD and the slope of the calibration curve, b, by

LOD = 3 Sa / b

Limit of Quantitation (LOQ)

The LOQ is the concentration that can be quantitated reliably with a specified level of accuracy and precision.The LOQ represent the concentration of analyte that would yield a signal-to-noise ratio of 10.

LOQ = 10 Sa/b

Where, Sa is the standard deviation of the peak area ratio of analyte to IS (6 injections) of the drugs and b is slope of the corresponding calibration curve.

Results and Discussion

Optimization of chromatographic conditions

The chromatographic conditions were optimized by different means i.e. using different column,different mobile phase, different flow rate, different detection wavelength and different diluents for standard drug are summarized in Table 1&2 and the optimised chromatogram (Figure 1) is shown.

Figure 1: Chromatogram of Imatinib standard.

Appreciable results were obtained by using mobile phase consisting of Potassium dihydrogen phosphate buffer: acetonitrile (70:30) on phenomenex C-18, (250mm*4.6mm) column with wavelength of detection of 268 nm. Flow rate was fixed at 0.8 ml/ min with a run time of 10 min.

Table 1: Optimization of chromatographic conditions of Imatinib.

Trials

Column Used

Mobile Phase

Flow Rate

Wave

length

Observation

Result

 

1

Hiq Sil, C-18, Vsize

(250mm*4.6mm)

Potassium dihydrogen phosphate Buffer

: ACN(50:50)

 

1 ml/min

 

268nm

Peak tailing and negative peak was found.

Method Rejected

 

 

2

 

Hiq Sil, C-18,

V size (250mm*4.6mm)

 

Potassium dihydrogen phosphate Buffer: Methanol: ACN(50:30:20)

 

 

1 ml/min

 

 

268nm

 

Poor resolution. Peak fronting and a broad Peak were found.

 

Method Rejected

 

 

3

 

waters, C-18, (250mm*4.6mm)

 

Potassium dihydrogen phosphate Buffer

: ACN (60:40)

 

 

1 ml/min

 

 

268nm

 

Peak shape was not good and also a tailing was found.

 

Method Rejected

 

4

waters,C-18,

(250mm*4.6mm)

Potassium dihydrogen phosphate

buffer: ACN (65:35)

 

1 ml/min

 

268nm

Poor resolution and peak tailing was

also found.

Method

Rejected

 

5

waters, C-18, (250mm*4.6mm)

Potassium dihydrogen phosphate buffer: acetonitrile (70:30)

 

0.8 ml/min

 

268nm

Good resolution, theoretical plate count and less tailing factor

Method Accepted

Table 2: Linearity data of Imatinib.

Conc. AUC (n -=6)
0 0
5 343726
10 801625
15 1064970
25 1811846
35 2721573

A chromatogram of Imatinib standard was made to run by injecting the solution prepared in section 2.4 in to HPLC.

Specificity

Chromatogram obtained for the injection is shown figure 2 with Rt of 2.67 mins without the use of any internal standard.

Figure 2: Figure showing Specificity.

Linearity & Range

The calibration curve showed in figure 3 has good linearity in the range of 5–35 µg/ml (Table 3),for Imatinib (API) with correlation coefficient (r2) of 0.9963.

Figure 3: Calibration curve for Imatinib.

The slope and intercept of the calibration graph was calculated by using linear regression analysis.The regression equation of the calibration curve was: y =76594x- 24947;r2 = 0.996.A correlation coefficient of 0.996 suggests that the developed HPLC method had an excellent linearity over the investigated range.

Table 3: Precision data of Imatinib.

HPLC Injection Replicates of Imatinib Retention Time Area
Replicate – 1 2.64 789939
Replicate – 2 2.67 790996
Replicate – 3 2.68 809774
Replicate – 4 2.67 796107
Replicate – 5 2.69 821313
Average 2.67 801625.8

Precision Repeatability:

The precision of each method was ascertained separately from the peak areas & retention times obtained by actual determination of five replicates of a fixed amount of drug. Imatinib (API).The percent relative standard deviation was calculated for Imatinib are presented in the table 3 and Overall repeatability for Imatinib is shown in figure 4.

Figure 4: Overall repeatability for Imatinib.

Intra & inter day precision:

The intra & inter day variation of the method was carried out & the high values of mean assay & low values of standard deviation & % RSD (% RSD < 2%) within a day & day to day variations for Imatinib revealed that the proposed method is precise and shown in table 4.

Table 4: Data for intra-day assay & inter-day assay.

Conc. of Imatinib (API) (µg/ml) Observed Conc. of Imatinib (µg/ml) by the
proposed method
Intra-Day Inter-Day
Mean (n=9) % RSD Mean (n=6) % RSD
5 4.81 0.96 10.03 0.592193
10 20.04 0.4 30.03 1.074387
15 39.97 0.33 39.95 0.452849

Accuracy: Recovery study

The recovery of the method, determined by adding a previously analyzed test solution with additional drug standard solution at three levels of concentration, was 99.99- 100.46 %.The values of recovery (%) and RSD (%) listed in Table 5 indicate the method is accurate.

Table 5: Data for Recovery of Imatinib.

Sample ID Concentra tion (µg/ml) %Recovery of Statistical Analysis
Pure drug Formulation Pure drug
S1 : 80 % 8 10 99.63 Mean= 99.67667%
S2 : 80 % 8 10 99.92 S.D. = 0.223681
S3 : 80 % 8 10 99.48 % R.S.D.= 0.224407
S4 : 100 % 10 10 99.19 Mean= 99.19%
S5 : 100 % 10 10 99.25 S.D. = 0.06
S6 : 100 % 10 10 99.13 % R.S.D.= 0.06049
S7 : 120 % 12 10 99.25 Mean= 99.49%
S8 : 120 % 12 10 99.54 S.D. = 0.219317
S9 : 120 % 12 10 99.68 % R.S.D. = 0.220441

Accuracy: The mean recovery was found to be 99.882% for Imatinib and shown in table 5.The limit for mean % recovery is 98-102% and as both the values are within the limit,hence it can be said that the proposed method was accurate.

Limit of detection and limit of quantification

The LOD was found to be 0.341µg/ml and LOQ was found to be 1.023 µg/ml for Imatinib which represents that sensitivity of the method is high.

System suitability

The system suitability parameters are shown in table 6.

Table 6: Data for system suitability parameters.

S. No. Parameter Limit Result
1 Resolution Rs > 2 9.15
2 Asymmetry T ≤ 2 Imatinib=0.12
3 Theoretical plate N > 2000 Imatinib=3246

Estimation of Imitanib in Capsule Dosage Form

Assay was performed by using the regression equation (y =76594x-24947;r2 =0.996) obtained from the standard curve of Imitanib API.Results obtained are given in table 7 and represented as chromatogram in figure 5.The amount of drugs in omef capsule was found to be 99.10 (±0.498) mg/tab for Imatinib.

Figure 5: Chromatogram of Assay of Imatinib Capsule.

Table 7: Data for Assay of Imatinib drug in marketed capsule.

Brand name of Labelled amount Mean (±SD) amount (mg) found by %
Capsules of Drug (mg) the proposed method (n=6) RSD
Omef 20 99.10 (±0.498) 0.494

Conclusion

A New RP-HPLC method indicating assay of Imatinib in bulk and in pharmaceutical dosage form is established.This method is simple, reliable, linear, accurate, sensitive and reproducible as well as cost effective for the effective quantitative analysis of imatinib in bulk and capsule formulations.The method was completely validated showing satisfactory data for all the method validation parameters tested and method is free from interference of the other active ingredients and additives used in the formulations.Therefore the method is suitable for use of the routine quality control analysis of imatinib in API or in pharmaceutical dosage forms.

References

1. Demetri GD, Mehren MV, Blanke CD, Van den Abbeele AD, Eissenberg B, Roberts PJ, et al. Efficacy and safety of Imatinib mesylate in advanced gastrointestinal stromal tumors. N Engl J Med. 2002; 347: 472-480.

2. Alkharfy KM, Khan RM, Al Asmari M, Alhadeyah BH, Ahmad. A Quantitative determination of Imatinib stability under various stress conditions. J Pharm Bioallied Sci. 2013; 5: 49-52.

3. Bende G, Kollipara S, Sekar V, Saha R. UV spectrophotometric determination of Imatinib mesylate and its application in solubility studies. Pharmazie. 2008; 63: 641-645.

4. Bakshi M, Singh S. Development of validated stability indicating assay methods critical review. J Pharm Biomed Anal. 2002; 28: 1011-1040. 

5. Kuna AK, Kuna J K. RPHPLC method development and validation of imatinib mesylate in tablet dosage form. Int J Pharm Pharm Sci. 2011; 3: 162-165.

6. Medenica M, Janic B, Ivanoic D, Malenovic A. Experimental design in reverse phase high performance liquid chromatographic analysis of imatinib mesylate and its impurity. J Chromatogr A. 2004; 1031: 243-248.

7. Rosasco, Maria A, Moyano, Pizzorno, Maria T, Segall,Adriana I. Validation of an HPLC method for the determination of Imatinib mesylate in pharmaceutical dosage. J Liq Chromatogr and related Tech. 2005; 28: 3283-3292.

8. Padala RR, Rajaval P, Venkateswararao P. Method development and validation of imatinib in capsule dosage form by using RP-HPLC method. International Journal of Research in Pharmaceutical and nano sciences. 2013; 2: 689-696.

9. Satyanarayana G, Ramesh E, Kumar JP, Rao HK, Sridhar B, Nagaraju P et al. Development and validation of new reversed phase high performance liquid chromatography method for the estimation of Imatinib in bulk and pharmaceutical dosage forms. International Journal of Research in Pharmaceutical and Biomedical Sciences. 2010; 1: 6-9.

10. Sandhya P, Vishnu Priya P, Shymala, Anjali Devi N, Sharma JVC. Method development and validation of imatinib mesylate in pharmaceutical dosage form by RP-HPLC. World Journal of Pharmacy and Pharmaceutical Sciences. 2013; 3: 682-688.

11. Aleti P, Venisetty RK, Kamarapu SK. Development of RP HPLC method for the analysis of imatinib mesylate using PDA detector and its application in the evaluation of marketed preparation. Int J Pharm Bio Sci. 2011; 1: 14-18.

12. Teoh M, Narayanan P, Moo KS, Radhakrissman S, Pillappan R, Bukhari NI, et al. HPLC determination of imatinib in plasma and tissues after multiple oral dose administration to mice. Pak J Pharm Sci. 2010; 23: 35-41.

13. Titier K, Picard S, Ducint D, Teihet E, Moore N, Berthaud P et al. Quantification of imatinib in human plasma by high performance liquid chromatography tadem mass spectrometry. Ther Drug Monit. 2005; 27: 634-40.

14. ICH.Validation of analytical procedures: Methodology harmonized tripartiate guideline prepared within the international conference on harmonization of technical requirements for the registration of pharmaceuticals for human use. ICH-Q2B: Geneva; 1996.

Other Articles

Article Image 1

Removal of Acid Beverage Flocs in Crystal Sugar by Adsorption Column Chromatography: Preliminary Study with Adsorbent Resin

The sugar industry has been facing some problems with white sugar intended for soft drinks production. Some studies have been demonstrated the sugar could be source of insoluble substances technically called by Acid Beverage Flocs (ABF). It appears during the beverage production; however, the flocs are easily disappearing with shaking. This may be related to raw material the sugarcane quality and, consequently, with the efficiency of the juice clarification process, since if there is no adequate juice treatment. The sugarcane commonly has lower quality and higher intensity of impurities, such as proteins, wax, minerals and colloids. The sugar industries are implementing solutions to reduce the amount of impurities through using technologies like ionic resins to removal of color, odor or undesirable taste. It could transform raw sugar in refined or white sugars. The aim of this work was evaluate the efficiency of purification of sugar solution with chromatographic adsorption system. We observed that the color reduction was around 69% and an intensive reduction in turbidity of sugar aqueous solution. There were variation on sucrose and increase on reducing sugars (glucose and fructose) contents. It is possible to conclude that the process could be efficient to reduce the color (MAU color), turbidity and ABF into sugar solutions.

Roberta Bergamin Lima and Claudio Lima de Aguiar*


Article Image 1

A Study to Evaluate Aflatoxin Contamination in Food from Gauteng Province

Aflatoxins (AFs) and their producers have been found to be problematic and is still an issue. This is due to their role in influencing food quality, health and economy loss. In maintaining good food quality, minimize AFs occurrence and its deleterious effects, it is essential that issues relating to mycotoxin production and contamination are monitored, and addressed on a regular basis. This will proffer suggestions for immediate actions to be taken, in order to achieve and maintain regulatory limit before situations are out of control. Hence the case study to evaluate the concentration of AFs in food, by way of thin layer chromatography and high performance liquid chromatography was attempted. This is to monitor concentrations of the emerging mycotoxin in relation to international regulatory limits. Concentrations levels ranging 0.06 - 77.97 ppb was observed, (mean: 16.8 ppb) and measured out in the food samples of study, exceeding the ISO limits (0.05 ppb). This calls for proper monitoring of pre and postproduction stages in food, by employing easy to use chromatographic techniques as a routine process to screen and ascertain concentrations do not exceed regulatory limits.

Henry E Iheanacho¹,²,³*


Article Image 1

Investigation of Naturally Occurring Fumonisin B1 and Glycated Fumonisin B1 in Korean Feedstuffs

Mycotoxins are potentially dangerous contaminants of livestock feeds. In this study, we measured the levels of fumonisin B1 and glycated fumonisin B1 in feedstuffs and then investigated the ability of the extrusion heating regimen to convert the most prevalent mycotoxin contaminant, fumonisin B1 , to a less toxic glycated form. All feed samples were analyzed with fully validated methods. All measured concentrations of fumonisin B1 were below harmful thresholds, including European Union-recommended levels or US Food and Drug Administration action levels.

Because fumonisin B1 was highly contaminated mycotoxin in our present investigation and fumonisin B1 has been shown to be less toxic following Maillard type reaction with reducing sugar, we examined the formation of fumonisin B1 derivatives by Maillard reaction under extrusion process conditions. We employed a variety of tandem mass spectrometric methodologies to selectively detect fumonisin B1 derivatives and to elucidate their structures partially. We found that compounds of m/z 736 were more likely artifacts or side reaction products rather than glycation products. N-(carboxymethyl) fumonisin B1 of m/z 780 and other major glycation products of m/z 794 and 810 were not detected, and only negligible amounts of methylene fumonisin B1 was found in 10 extruded feed samples. Therefore, either the tested extrusion conditions did not induce fumonisin B1 glycation or the glycation products simply could not be detected by the method employed in this study.

Jongsung Ahn¹, Hyenjong Kim¹, and Kwang-Yeop Jahng²*


Article Image 1

Determination of HMF in Some Instant Foods and Its Biodegradation by Some Lactic Acid Bacteria in Medium and Food

The aims of this study were to determine HMF level in some foods and to reduce their level by biodegradation. Coffee whitener, cappuccino, hot chocolate, instant coffee, instant powdered mix (coffee, coffee whitener and sugar) and aromatized cocoa were subjected to HMF analysis in the scope of this study. The HMF determination of samples was carried out using a high performance liquid chromatography. Color properties of the samples were determined according to standard procedures and their relations with HMF content were also investigated. The biological degradation of HMF with lactic acid bacteria was also studied with broth media and a model food system (reconstituted milk).The average HMF levels were 12.59 mg/kg for coffee whitener, 572.49 mg/ kg for cappuccino, 660.29 mg/kg for hot chocolate, 1804.91 mg/kg for instant coffee, 871.56 mg/kg for instant powdered mix and 980.94 mg/kg for aromatized cocoa. The addition of lactic acid cultures reduced the HMF content of the samples. HMF levels of broth media and model food inoculated with lactic acid bacteria (L. lactis, L.bulgaricus, L. cremoris) decreased about 25 % as a result of HMF biodegradation.

Rasim Alper Oral¹, Mahmut Dogan², Kemal Sarıoglu², Ahmed Kayacıer¹, and Osman Sagdic³*


Article Image 1

An Integrated Analysis of the Musa Paradisiaca Peel, Using UHPLC-ESI, FT IR and Confocal Microscopy Techniques

When the banana (Musa paradisiaca) peel was subject of enzymatic hydrolysis with cellulase and hemicellulase, only glucose was obtained. Images from banana peel, using confocal laser scanning microscopy, demonstrate that the cellulose is the main structural compound. Fatty acids, phenolic and other compounds were detected from the organic residual extract, and characterized by means of NMR, FT-IR and UHPLC-MS techniques. The presences of these compounds were corroborated through a steam distillation. Under this condition, banana peel could have potential applications in the food field, where could be used to improve some procedures such as the obtaining of banana vinegar.

Miriam Alejandra González Corona¹, Mayra Beatriz Gómez-Patiño², María de Jesús Perea Flores², Luis Alberto Moreno Ruiz², Blanca Margarita Berdeja Martinez¹, and Daniel Arrieta-Baez²*


Article Image 1

Structural Characterization of Bamboo Lignin Isolated With Formic Acid and Alkaline Peroxide by Gel Permeation Chromatography and Pyrolysis Gas Chromatography Mass Spectrometry

Fractionation is an effective technology to maximize the utilization of lignocelluloses for the production of chemicals and materials. In this case, bamboo was subjected to a two-step fractionation process based on the concept of biorefinery: (a) formic acid treatment at boiling point under atmospheric pressure for 2 h, and (b) post treatment with alkaline hydrogen peroxide solution containing 1% NaOH and 1% H2 O2 at 80 ºC. The combination of formic acid delignification and alkaline hydrogen peroxide degradation achieved an effective removal of both lignin (delignification rate 94.9%) and hemicelluloses (removal rate 87.4%) from bamboo, producing cellulose rich pulp, formic acid lignin and sugars. To investigate the structural modification of lignin during the fractionation process, the residual lignin in the treated samples was isolated and characterized with multiply techniques including gel permeation chromatography, pyrolysis gas chromatography mass spectrometry, Fourier-transform infrared spectroscopy, etc. The relative ratio of S/G was 1.63 for bamboo milled wood lignin (L1), whereas the lignin isolated from the formic acid treated cellulose-rich fraction (L2) presented a chromatograph similar to that of L1 but had a lower S/G ratio of 1.28. This indicated that a preferential removal of S units during the formic acid fractionation process. In addition, alkaline hydrogen peroxide treatment resulted in more removal of S units, as indicated by a lower S/G ratio of 0.71.

Xun Li¹, Chang-Zhou Chen¹, and Ming-Fei Li¹*

 


Article Image 1

Volatile Compound Profiles by HS GC MS for the Evaluation of Postharvest Conditions of a Peach Cultivar

Volatile Organic Compounds (VOCs) profile of foods obtained by Gas Chromatography/Mass Spectrometry (GC/MS) can be considered a potent tool of food products quality changes occurring as a result of different processing, such as ripening and deterioration. The aim of the present study was the evaluation of volatiles profiles of peaches (cv Springcrest) during their storage in conditions similar to those of long distance transport that normally these products undergo before being placed on market. We investigated control sample (no stored fruit) and peaches stored in cardboard boxes wrapped in heat-sealed HD polythene bags, both in normal and modified atmosphere (0% and 23% CO2 ) after 1 and 8 days of storage at 4°C. GC/MS analysis of these samples allowed the identification of a total of 115 VOCs.

The comparison of the VOCs profile of the three peach samples (control, normal atmosphere and 23% CO2 ) shows that fruits packaged in normal atmosphere released a greater amount of esters of medium chain fatty acids, such as ethyl nonanoate and ethyl dodecanoate. On the other hand, fruits stored in normal atmosphere and modified atmosphere after 8 days of storage (increased concentration of CO2 in packs) released a greater amount of esters of long chain fatty acids, such as ethyl hexadecanoate.

Livia Malorni¹, Antonella Martignetti¹, and Rosaria Cozzolino¹*


Article Image 1

Current Trends in Lignocellulosic Analysis with Chromatography

The conversion of lignocellulosic biomass into biofuel and biomaterial is promising for the substitution of fossil resources in energy and material applications. Given the complexity of plant cell wall, the main challenge is to obtain lignocelluloses with high yield and purity. For a better understanding of lignocellulosic biomass, chromatography stands out as a powerful separation method that can support the lab directed research and pilot scale production of biomaterial and biochemical. This paper provides a review on the characterization of cellulose, hemicellulose and lignin along with their derivatives and decomposed sugar monomers, in particular their isolation and purificationmethods using various specific types of chromatography. Methods with various specific types of chromatography. This review also summarizes different chromatographic methods for obtaining the molecular weights of cellulose, hemicellulose and lignin that have been used in recent years, and highlights future opportunities for the application of those biopolymers.

Fengbo Sun¹ and Qining Sun²*


Article Image 1

Current Status of Two-Dimensional Gel Electrophoresis and Multi-Dimensional Liquid Chromatography as Proteomic Separation Techniques

Proteomics is very important component in the era of post-genomics because it can address functions of genes and some important non-gene-determined biological issues such as Post Translational Modifications (PTMs), splicing, translocation, and spatial structure. Proteome is very complex, including multiple parameters such as kind of proteins, copy number of each protein, PTMs, isoforms, spatial structure of each protein, protein-protein interaction, and protein-other molecule interaction, etc. Moreover, proteome is dynamic, and alters with different conditions such as different physiological processes, different pathological processes, and different disease status.

Xianquan Zhan¹,²,³,⁴*


Article Image 1

Chromatography & Separations

I started off my career as a biotechnologist in the early eighties, of the last century, using mainly size exclusion chromatography to identify biologically active substances, from ovine broncho alveolar macrophages co-cultured with lymphocytes, from the lungs of young sheep infected with Jaagsiekte Retrovirus (JSRV)

MS Myer1*